Code 30: setting of calendar – Yokogawa GD40 Gas Density Detector User Manual
Page 83

<5. DENSITY / PARAMETER SETTING>
5-28
IM 11T3E1-01E
CODE 30: Setting of Calendar
This setting defines the date and time.
(a) Setting of year/month/day [*Y_M_D]
Configurable range : 00.01.01 to 99.12.31
(b) Setting of hour/minute [*H_M]
Configurable range : 00.00 to 23.59
30
*CODE
*SERVC
XX.XX.XX
*Y_M_D
XX.XX
*H_M
F0522.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
If you enter a value that exceeds the limits, the display shows [*OVER].
Press the [YES] or [NO] key to re-set the value.
[ENT]
[ENT]
[YES]
[ENT]
*
3
*
3
*
2
*
1
Figure 5.21 Setting of Calendar
CODE 31: Showing/Hiding of Negative Measured Values [*MINUS]
This setting determines whether negative measured values (-) should be shown or hidden.
・ Show: 0
・ Hide: 1
[ENT]
*CODE
31
*SERVC
X
*MINUS
[ENT]
[YES]
F0523.ai
*1: See "Figure 5.3 Process at the Setting Level" on page 5-10.
*2: Enter the code number.
*3: Set a value using the [ ], [ ] and [ENT] keys.
*
3
*
2
*
1
Figure 5.22 Showing/Hiding of Negative Measured Values