beautypg.com

Yokogawa GD40 Gas Density Detector User Manual

Page 112

background image

<6. CALORIE / PARAMETER SETTING>

6-23

IM 11T3E1-01E

CODE 13: Selection for Automatic Calibration [*AUTO.C]

This setting determines whether or not automatic calibration is carried out. If automatic calibration

is to be carried out, you must also configure CODE 15 at the service level.

・ Disable:

0

・ Enable:

1

.

[ENT]

*CODE

13

*SERVC

X

*AUTO.C

[ENT]

[YES]

F0615

*

1

*

2

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

Figure 6.15 Selection for Automatic Calibration

CODE 14: Setting for Remote Semi-automatic Calibration [*REMOT]

This setting determines whether or not semi-automatic calibration is carried out remotely.

If remote semi-automatic calibration is to be carried out, you must wire the contact input.

You must also configure CODE 15 at the service level.

・ Disable:

0

・ Enable:

1

[ENT]

*CODE

14

*SERVC

X

*REMOT

[ENT]

[YES]

F0616.ai

*

1

*

2

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

*

3

Figure 6.16 Setting for Remote Semi-automatic Calibration

This manual is related to the following products: