11.2 Inspection In Case of Failure

11.2.1 Inspecting the Analyzer in an Alarm Status

The GD402 analyzer outputs an ALARM signal through terminals 16 and 17 if a failure is de-

tected by the self-diagnostic function during measurement. In addition, the lamp on

the operation panel comes on and the message field shows an alarm number. The analyzer out-

puts no contact signal, however, if the alarm number is [ALM.09]. If a failure (alarm) is detected,

take the corrective action given in Table 11.1.

Table 11.1 Corrective Actions when an Alarm Status Is Detected

Message

Shown

Failure Type

Diagnostic Conditions

Corrective Actions

Message via

Communication

ALM.01

High/low limit alarm

against physical

density

Physical density ≤ Low-limit-of-

physica-density alarm setpoint

Physical density ≥ High-limit-of-

physical-density

Change the high/low

limits.

OUT OF DENSITY

ALM.02

High/low limit alarm

against compensated

density

Compensated density ≤ Low-limit-of-

compensated-density alarm setpoint

Compensated density ≥ Hgh-limit-of-

compensated-density alarm setpoint

Change the high/low

limits.

OUT OF COMP

DENSITY

ALM.03

High/low limit alarm

against specific gravity

Specific gravity ≤ Low-limit-of-specific

gravity alarm setpoint

Specific gravity ≥ High-limit-of-specific

gravity alarm setpoint

Change the high/low

limits.

OUT OF SP GR

ALM.04

High/low limit alarm

against calorific value

Calorific value ≤ Low-limit-of-calorific-

value alarm setpoint

Calorific value ≥ High-limit-of-calorific-

value alarm setpoint

Change the high/low

limits.

OUT OF CALORY

ALM.05

High/low limit alarm

against molecular

weight

Molecular weight ≤ Low-limit-of-

molecular weight alarm setpoint

Molecular weight ≥ High-limit-of-

molecular weight alarm setpoint

Change the high/low

limits.

OUT OF

MOLECULE

ALM.06

High/low limit alarm

against concentration

Concentration ≤ Low-limit-of-

concentration alarm setpoint

Concentration ≥ High-limit-of-

concentration alarm setpoint

Change the high/low

limits.

OUT OF CONC

ALM.07

Error in pressure input

range

Zero point -3% ≥ pressure

Span point +5% ≤ pressure

Pressure ≤ 0.1 kPa

Check the sample-gas

pressure and the

pressure range.

OUT OF GAS

PRESS

ALM.08

Abnormal sample gas

temperature

-25 °C ≤ sample gas temperature

≤ 80 °C

Use the sample gas

within the tolerance

limits.

OUT OF GAS

TEMP

ALM.09

Battery failure

Contact Yokogawa

service personnel.

LOW BATTERY

ALM.10

Error in calibration

(zero and span points)

-0.3000 ≤ zero calibration value

≤ 0.3000

0.50000 ≤ span calibration value

≤ 1.5000

Redo calibration.

ILLEGAL CALIB