beautypg.com

Code 02: setting of hold in the event of errors, Code 03: setting of hold of hydrogen purity range – Yokogawa GD40 Gas Density Detector User Manual

Page 139

background image

<7. HYDROGEN PURITY / PARAMETER SETTING>

7-16

IM 11T3E1-01E

CODE 02: Setting of Hold in the Event of Errors

This setting determines whether or not the output signal is held if an error occurs. You can select

either the value immediately prior to the error or a preset value as the value to be held.

(a) Hold enable/disable [*E_HLD]
・Disable:

0

・Enable:

1

・Enable (preset value):

2

(b) Setting of preset hold value [*PR.SET]
This item applies if item (a) above is set to "2".
Configurable range: -10.0 to 110.0 (percent of analog output)

[ENT]

02

*CODE

*SERVC

X

*E_HLD

XXX.X

*PR.SET

[YES]

F0709.ai

*

1

*

2

*

4

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ] and [ENT] keys.

*4: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*5: Setting 0 or 1 returns to the service level with the indication [*SERVC].

[ENT]

[ENT]

*

3

*

5

Figure 7.9 Setting of Hold in the Event of Error

CODE 03: Setting of Hold of Hydrogen Purity Range

This setting enables or disables the function for holding the output in the hydrogen purity range

(hydrogen purity/replacement (substituents-concentration) meters mode only). You can select

either the value immediately prior to the hold or a preset value as the value to be held.

(a) Hold enable/disable [*H_HLD]
• Disable:

0

• Enable:

1

• Enable (preset value): 2

This manual is related to the following products: