beautypg.com

2 manual calibration, Manual calibration -5 – Yokogawa GD40 Gas Density Detector User Manual

Page 158

background image

<8. DENSITY / CALIBRATION PROCEDURE>

8-5

IM 11T3E1-01E

8.3.2

Manual Calibration

In this mode of calibration, you visually make sure the reading has settled, and then confirm the

reading manually (by pressing the [ENT] key). There are two ways to switch between the zero,

span and sample gases: by using manually-operated valves or by using solenoid valves. When

switching between the gases using solenoid valves, follow the instructions on valve installation in

Subsection 8.3.1.

[NO]

SEM.CAL

[NO]

VALVE

MAN.CAL

X.XXXX

Z_DNS

X.XXXX

WAIT

F0805.ai

[YES]

*

1

ALM.10

ZERO

[YES]

Measurement mode

*

2

*

5

[NO]

X.XXXX

S_DNS

[YES]

*

3

*

6

SPAN

*

4

X.XXXX

CAL.SET

X.XXXX

CAL.SET

[ENT]

[ENT]

[ENT]

[ENT]

[YES] or [NO]

*

3

*

4

[NO]

*1: See "Figure 5.1 Modes of Operational Level" on page 5-8.

*2: If a calibration data error occurs, the display shows [ALM.10].

Pressing the [YES] or [NO] key returns the display to the [MAN.CAL].

*3: Set a desired value by using the [ ], [ ] and [ENT] keys.

*4: When the reading settles, press the [ENT] key to confirm your calibration.

This parameter does not accept input with the [ ] and [ ] keys.

If you press the [MODE] key (for ESCAPE) after this parameter appears, the display shows [WAIT].

The display always shows the corrected density.

*5: Press the [YES] key to go into the measurement mode status after a given stabilization time elapses: Press the [NO]

key to go into the state immediately.

*6: In case of pressing the [NO] key after skipping zero calibration (by pressing [NO] key in the ZERO status), the

analyzer returns to [MAN.CAL].

In case of pressing the [NO] key after finishing zero calibration, the analyzer goes to [WAIT] status.

Zero-point density display

Span-point density display

Figure 8.4 Manual Calibration Mode

This manual is related to the following products: