<9. CALORIE / CALIBRATION PROCEDURE>

9-3

IM 11T3E1-01E

9.2.2

Setting Calibration Data

To set calibration data, press the [*] key in the measurement mode, (when password is selected,

enter the password (XXX), ) and select the parameter [*CAL.DT]. See Figure 6.4 for the process

at the setting level.

(a) Zero-point calorific value [*Z_CAL]

Configurable range : 000.000 to 133.000 (MJ/m

3

)

0.0000 to 3.5000 (kBTU/ft

3

)

(b) Zero-point density [*Z_DNS]

Configurable range : 0.0000 to 6.0000 (kg/m

3

)

0.00000 to 0.40000 (lb/ft

3

)

(c) Span-point calorific value [*S_CAL]

Configurable range : 000.000 to 133.000 (MJ/m

3

)

0.0000 to 3.5000 (kBTU/ft

3

)

(d) Span-point density [*S_DNS]

Configurable range : 0.0000 to 6.0000 (kg/m

3

)

0.00000 to 0.40000 (lb/ft

3

)

(e) Output hold value during calibration [*C_HLD]
• Disable:

0

• Enable:

1

• Enable (preset value): 2
(f) Setting of preset value [*PR.SET]

Configurable rang :

-10.0 to 110.0

(Applies only if item (e) above is set to “2”.)