Yokogawa GD40 Gas Density Detector User Manual
Page 102
![background image](/manuals/360159/102/background.png)
<6. CALORIE / PARAMETER SETTING>
6-13
IM 11T3E1-01E
1. Analog Output Setting Mode [ *RANGE ]
The analog output has output 1 and output 2. Only output 1 can be used for communication
purposes.
Set a range appropriate for 4-20 mA DC output signals. The range has two setpoints: the lower
limit (zero point) of the range corresponding to the minimum (0%) of a given output signal and the
upper limit (span point) of the range corresponding to the maximum (100%) of the given output
signal.
(1) Output 1
[*OUT1]
(a) Select a parameter to be output.
・ Physical density [*DENS]
・ Compensated density [*C_DNS]
・ Calorific value [*CALRY]
・ Gas temperature [*TEMP]
・ Gas pressure [*PRESS]
(b) Set the zero and span points. (This item is not applied if temperature or pressure are cho-
sen as the parameter in item (a) above.)
・ Physical density [*Z_DNS], [*S_DNS]
Configure range:
00.0000 to 60.0000 (kg/m
3
),
0.00000 to 4.00000 (lb/ft
3
)
・ Compensated density [*Z_CP.D], [*S_CP.D]
Configure range:
0.0000 to 6.0000 (kg/m
3
),
0.00000 to 0.40000 (lb/ft
3
)
・ Calorific value [*Z_CAL], [*S_CAL]
Configure range:
000.000 to 133.000(MJ/m
3
),
0.0000 to 3.5000 (kBTU/ft
3
)
(2) Output 2
[*OUT2]
The same parameters as those of item (1) apply.