beautypg.com

2 parameter setting, 1 parameter setting at operation level, Parameter setting -6 5.2.1 – Yokogawa GD40 Gas Density Detector User Manual

Page 61: Parameter setting at operation level -6

background image

<5. DENSITY / PARAMETER SETTING>

5-6

IM 11T3E1-01E

Table 5.4 Setting Parameters at Service Level

Code

No.

Item

Display

Data to be Set

(or Conditions)

Remarks

20 Selection of pressure unit

*PRES.U

· kPa : 0

· MPa : 1

· psi : 2

See Figure 5.17 for key operation.

21 Selection of density unit

*DENS.U

· kg/m

3

: 0

· lb/ft

3

: 1

See Figure 5.18 for key operation.

22 Selection of calorific value unit

*CAL.U

· MJ/m

3

: 0

· kBTU/ft

3

: 1

See Figure 5.19 for key operation.

23 Selection of temperature unit

*TEMP.U

· °C : 0

· °F : 1

See Figure 5.20 for key operation.

30 Setting of calendar

Setting of year/month/day

Setting of hour/minute

*Y_M_D

*H_M

00.01.01 to 99.12.31

00.00 to 23.59

See Figure 5.21 for key operation.

For leap years, the setting can

cope with ‘February 29’.

31 Showing/hiding of negative measured

values (-)

*MINUS

· Show : 0

· Hide : 1

See Figure 5.22 for key operation.

40 Indication of calibration coefficients

(read-only)

Zero

Span

*C_K_Z

*C_K_S

SX.XXXX

X.XXXX

See Figure 5.23 for key operation.

S of SX.XXXX means Marks such

as a plus or the minus.

41 Indication of oscillation frequency

(read-only)

Frequency F2

Frequency F4

Frequency ratio

*F2.KHZ

*F4.KHZ

*F2/F4

X.XXXXX

X.XXXXX

X.XXXXX

See Figure 5.24 for key operation.

42 Indication of software version

(read-only)

*REV

X.XX

See Figure 5.25 for key operation.

43 Setting of high-resolution mode

*S_CYC

· Normal :

0

· High-resolution : 1

See Figure 5.26 for key operation.

44 Setting of password

*PASS

0.0.0 to 9.9.9

See Figure 5.27 for key operation.

45 Selecting Battery alarm detection

*BAT

· Non detection : 0

· Detection : 1

See Figure 5.28 for key operation.

50 Selecting the instruments

*MODEL

· Density meter : 0

· Calorie meter : 1

· Hydrogen purity, replacement

meter : 2

See Figure 5.29 for key operation.

82 Setting Detector Constant

See Figure 5.30 for key operation.

5.2

Parameter Setting

[Aborting the Setting Operation]
Press the [MODE] key. Normally, the [MODE] key is used to move from the measurement

() mode to a mode at the operation level. It is also used to return to the measure-

ment mode from other modes.

5.2.1

Parameter Setting at Operation Level

To move to the operation level, press the [MODE] key and then enter the password (XXX). Pass-

word is not selected in first condition. When password is necessary, refer to Section 4.2 “Setting

Lists” and selected password what you need.

This manual is related to the following products: