<6. CALORIE / PARAMETER SETTING>

6-19

IM 11T3E1-01E

CODE 05: Setting of Contact Outputs Status [*CNTCT]

This setting defines the status of the contact outputs.

Configurable range: 00 to 15

[ENT]

05

*CODE

*SERVC

XX

*CNTCT

[ENT]

[YES]

F0611.ai

*

1

*

2

*

3

*1: See "Figure 6.4 Process at the Setting Level" on page 6-12.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

Figure 6.11 Setting of Contact Outputs Status

Each contact output takes either of the following two status (in table); depending on the value

you set.

Value

ZERO contact

SPAN contact MAINT contact

ALM contact

0

NO

NO

NO

NO

1

NO

NO

NO

NC

2

NO

NO

NC

NO

3

NO

NO

NC

NC

4

NO

NC

NO

NO

5

NO

NC

NO

NC

6

NO

NC

NC

NO

7

NO

NC

NC

NC

8

NC

NO

NO

NO

9

NC

NO

NO

NC

10

NC

NO

NC

NO

11

NC

NO

NC

NC

12

NC

NC

NO

NO

13

NC

NC

NO

NC

14

NC

NC

NC

NO

15

NC

NC

NC

NC

NO: means “normally open”

NC: means “normally closed”