beautypg.com

Code 12: setting of gas pressure – Yokogawa GD40 Gas Density Detector User Manual

Page 145

background image

<7. HYDROGEN PURITY / PARAMETER SETTING>

7-22

IM 11T3E1-01E

CODE 12: Setting of Gas Pressure

This setting defines the zero and span points of the pressure transmitter's measurement range.

(a) Setting of zero-point gas pressure

[

*Z_PRS

]

Configurable range: 0.01 to 999.99 (for kPa),

0.0001 to 9.9999 (for MPa),

0.015 to 99.999 (for psi )

(b) Setting of span-point gas pressure

[

*S_PRS

]

Configurable range: 0.10 to 999.99 (for kPa),

0.0001 to 9.9999 (for MPa),

0.015 to 99.999 (for psi )

The zero and span points set here are also used for the analog output.

[ENT]

*CODE

12

*SERVC

XXX.XX

*Z_PRS

XXX.XX

*S_PRS

[ENT]

[ENT]

[YES]

F0715.ai

*

1

*

2

*

3

*

4

*1: See "Figure 7.4 Process at the Setting Level" on page 7-10.

*2: Enter the code number.

*3: Set a value using the [ ], [ ] and [ENT] keys.

If you enter a value that exceeds the limits, the display shows [*OVER].

Press the [YES] or [NO] key to re-set the value.

*4: The position of the decimal point changes depending on the setting of CODE 20.

If the code is set to "0" (kPa), the reading is XXX.XX;

if the code is set to "1" (MPa), the reading is X.XXXX;

if the code is set to "2" (psi), the reading is XX.XXX.

*

3

*

4

Figure 7.15 Setting of Gas Pressure

This manual is related to the following products: