8. Detailed Data Setting
8. Detailed Data Setting
8.1 Current Output Setting
This section describes setting of the analog output range. Table 8.1 shows the parameter
codes for each setting item and set values.
Table 8.1 Display Items
T8.1.1.EPS
Set item
Parameter code Set value
Current output
C01
0: Oxygen concentration
1: Humidity
2: Mixing ratio
Output mode
C03
0: Linear
1: Logarithm
Min. oxygen concentration
C11
Oxygen concentration reading corresponding to 4 mA
Max. oxygen concentration
C12
Oxygen concentration reading corresponding to 20 mA
Min. humidity
C13
Humidity reading corresponding to 4 mA
Max. humidity
C14
Humidity reading corresponding to 20 mA
Min. mixing ratio
C15
Mixing ratio at 4 mA
Max. mixing ratio
C16
Mixing ratio at 20 mA
Output damping coefficient
C30
0 to 255 seconds
CAUTION
When you select logarithmic mode in Section 8.1.3, “Output Mode,” later in this
manual, the oxygen concentration, humidity reading, and mixing ratio remain constant at
0.1% O
2
, 0.1% H
2
O and 0.01 kg/kg respectively.