Startup, Operation of infrared switch -3 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 74
IM 11M12A01-05E
7-1
7. Startup
7. Startup
The following describes the minimum operating requirements — from supplying power
to the converter to analog output confirmation to manual calibration.
F7.0E.EPS
Place in normal operation
Check piping and
wiring connections
Set up valves
Supply power
Confirm converter
type setting
Select gas to be measured
Set output ranges
Check current loop
Check contact action
Calibrate analyzer
Set detailed data
Figure 7.1 Startup Procedure
System tuning by the HART communicator, refer to IM 11M12A01-51E ''HART
Communication Protocol''.