Err-01, Passno, G0 1 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 138: G3 0, Table 10.6 resetting
IM 11M12A01-05E
10-11
10. Other Functions
Table 10.6 Resetting
Switch operation Display Brief description
.
`
ENT
Err-01
If an error occurs, the error number and "------" are displayed
/------
alternately, as given on the left.
.
`
ENT
PASSno
Hold down the [ENT] key for at least three seconds.
.
`
ENT
0
000
Touch the [ENT] key again to switch to the password entry display.
.
`
ENT
1
000
Enter the password 1102.
Intermediate switch operations omitted.
.
`
ENT
1102
.
`
ENT
A
01
.
`
ENT
G
01
Change the parameter code to "G30."
.
`
ENT
G
0
1
.
`
ENT
G
3
0
.
`
ENT All the digits Touch the [ENT] key to execute resetting.
light up.
T10.5aE.EPS
The symbol ( ) indicates that the key are light characters indicate "flashing".
"/" indicate that the characters are displayed alternately.