4 setting purging – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 117
IM 11M12A01-05E
8-20
8.6.4
Setting Purging
Purging is to remove condensed water in the calibration gas pipe by supplying a span
calibration gas for a given length of time before warm-up of the detector. This prevents
cell breakage during calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after
the purge time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 1008 C or below upon power up and the
purge time is set in the range of 1 to 60 minutes.
F8.03.EPS
Displayed alternately
Figure 8.3 Display during Purging
8.6.4.1 Procedure
Use the parameter-code table below to set the purging time.
The allowable input ranges from 0 to 60 minutes.
Table 8.17 Parameter Code
Set item
Parameter code Set range Units
T8.17E.EPS
Purging time
F15
0 to 60
minutes
8.6.4.2 Default Value
When the analyzer is delivered, or if data are initialized, purging time is set to 0
minutes.