Alarm-related items in group d – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 142
IM 11M12A01-05E
10-15
10. Other Functions
Alarm-related Items in Group D
T10.6D.EPS
Code Item
Tuning
Engineering unit Default
setting
D01 Oxygen concentration, high-high
0 to 100
% O
2
100% O
2
alarm setpoint
D02 Oxygen concentration, high-limit
0 to 100
% O
2
100% O
2
alarm setpoint
D03 Oxygen concentration, low-limit
0 to 100
% O
2
0% O
2
alarm setpoint
D04 Oxygen concentration, low-low
0 to 100
% O
2
0% O
2
alarm setpoint
D05
Humidity, high-high alarm setpoint 0 to 100 %H
2
O 100% H
2
O
D06
Humidity, high-alarm setpoint 0 to 100 % H
2
O 100 % H
2
O
D07
Humidity, low-alarm setpoint 0 to 100 % H
2
O 0 % H
2
O
D08
Humidity, low-low alarm setpoint 0 to 100 % H
2
O 0 % H
2
O
D11
Mixing ratio, high-high alarm setpoint 0 to 1 kg/kg 1 kg/kg
D12
Mixing ratio, high-alarm setpoint 0 to 1 kg/kg 1 kg/kg
D13
Mixing ratio, low-alarm setpoint 0 to 1 kg/kg 0 kg/kg
D14
Mixing ratio, low-low alarm setpoint 0 to 1 kg/kg 0 kg/kg
D30 Oxygen concentration alarm
0 to 9.9
% O
2
0.1% O
2
hysteresis
D31
Humidity alarm hysteresis 0 to 9.9 % H
2
O 0.1 % H
2
O
D32
Mixing ratio alarm hysteresis 0 to 0.1 kg/kg 0.001 kg/kg
D33 Delayed alarm action
0 to 255
Seconds 3 seconds
D41 Oxygen concentration,
0: Not detected
Not detected
high-high alarm detection
1: Detection
D42 Oxygen concentration, high-limit
0: Not detected
Not detected
alarm detection
1: Detection
D43 Oxygen concentration, low-limit
0: Not detected
Not detected
alarm detection
1: Detection
D44 Oxygen concentration, low-low
0: Not detected
Not detected
alarm detection
1: Detection
D45
Humidity, high-high alarm detection 0: No detection
No detection
1: Detection
D46
Humidity, high-alarm detection 0: No detection
No detection
1: Detection
D47
Humidity, low-alarm detection 0: No detection
No detection
1: Detection
D48
Humidity, low-low alarm detection 0: No detection
No detection
1: Detection
D51
Mix ratio, high-high alarm 0: No detection
No detection
1: Detection
D52
Mix ratio, high-alarm detection 0: No detection
No detection
1: Detection
D53
Mix ratio, low-alarm
0: No detection
No detection
1: Detection
D54
Mix ratio, low-low alarm detection 0: No detection
No detection
1: Detection