16 contact-related items in group e – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 143
IM 11M12A01-05E
10-16
Contact-related Items in Group E
Code Item
Tuning
Engineering Default setting
unit
E01 Selection of input contact 1
0: Invalid
Invalid
1: Calibration gas pressure drop
2: Measurement range change
3: Calibration start
4: Detection of non-combusted gas
E02 Selection of input contact 2
0: Invalid
Invalid
1: Calibration gas pressure drop
2: Measurement range change
3: Calibration start
4: Detection of non-combusted gas
E03 Selecting action of input contact 1
0: Action with closed contact
Action with closed contact
1: Action with open contact
E04 Selecting action of input contact 2
0: Action with closed contact
Action with closed contact
1: Action with open contact
E10 Selecting action of output contact 1
0: Action with open contact (normally
Action with closed contact
energized)
1: Action with closed contact (normally
deenergized)
E20 Output contact 1 error
0: No action
No action
1: Action
E21 Output contact 1, high-high alarm
0: No action
No action
1: Action
E22 Output contact 1, high-limit alarm
0: No action
No action
1: Action
E23 Output contact 1, low-limit alarm
0: No action
No action
1: Action
E24 Output contact 1, low-low alarm
0: No action
No action
1: Action
E25 Output contact 1, during maintenance
0: No action
Action
1: Action
E26 Output contact 1, during calibration
0: No action
No action
1: Action
E27 Output contact 1, measurement range
0: No action
No action
change
1: Action
E28 Output contact 1, during warm up
0: No action
Action
1: Action
E29 Output contact 1, calibration gas
0: No action
No action
pressure decrease
1: Action
E32 Output contact 1, detection of non-
0: No action
No action
combusted gas
1: Action
T10.6E.EPS