8.2.2
Preference Order of Output Hold Values
The output hold value takes the following preference order:
During error occurrence
During calibration
During maintenance
During warm-up
Preference order (high)
8.2.2E.siki
For example, if the output current is set to 4 mA for “under maintenance”, and no
output-hold output for during calibration is preset, the output is held at 4 mA during the
“ under maintenance” display. However, the output hold is released at the time of
starting the calibration, and the output will be again held at 4 mA after completing the
calibration and when the output stabilization time elapses.
8.2.3
Output Hold Setting
Table 8.4 lists parameter codes with set values for individual set items.
Table 8.4 Parameter Codes for Output Holding
Set items
Parameter code
Set value
During warm-up
C04
0: 4 mA
1: 20 mA
2: Holds Set value
During maintenance C05
0: Without hold feature
1: Last measured value..
2: Holds set values.
During calibration
C06
0: Without hold feature
1: Last measured value.
2: Holds set values.
During error occurrence C07
0: Without hold feature
1: Last measured value.
2: Holds set values.
T8.5E.EPS
8.2.4
Default Values
When the analyzer is delivered or data are initialized, the output current settings are by
default as shown in Table 8.5.
Table 8.5 Output Current Default Values
Status
Output hold (min. and max. values)
During warm-up
4 mA
Under maintenance
Holds output at value just before maintenance started.
Under calibration or blow-back
Holds output at value just before starting calibration
or "blow-back."
On Error occurrence
Holds output at a preset value.
T8.6E.EPS
Preset value
4 mA
4 mA
4 mA
3.4 mA