7 current output setting, 1 analog output setting, Checking current loop -15 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 85: Checking contact i/o -16
IM 11M12A01-05E
7-12
7.7 Current Output Setting
7.7.1
Analog Output Setting
Select any one of the analog output settings — oxygen, humidity, and mixing ratio. If
the /HS option was specified at the time of purchase, the equipment is a humidity
analyzer. For other than this setting, the analyzer is an oxygen analyzer. If mixed
measurement is required, change the existing output setting as follows. Use parameter
code C01 for the setting (see Table 7.6). When the humidity analyzer is specified in the
above setting for the type of detector, the analog output will be set to “humidity” if data
initialization is performed.
Table 7.5.1 Analog Output Setting Procedure
T7501.EPS
.
.
.
.
.
.
The symbol ( ) indicates that the corresponding keys are being touched, and the light
characters indicate flashing.
.
.
.
.
.
.
.
.
A
01
C
01
0
1
1
1
C
01
Display after the password has been entered.
Change the parameter code to C01.
The current set value, in this case, 0: oxygen concentration,
is displayed.
Touch the [ ] key to set the number 1 for humidity setting.
Touch the [ENT] key. The numeric value will be flashing.
Touch the [ENT] key again to stop flashing.
Touch the [ENT] key to return to the parameter code
selection display.
ENT
ENT
ENT
ENT
ENT
ENT
ENT
.
Switch operation
Display
Description