Output-related Items in Group C
Code
Item
Tuning
Engineering Default setting
unit
C01
Analog output
0: Oxygen concentration
Humidity
1: Humidity
2: Mixed ratio
C03
Output mode
0: Linear
Linear
1: Logarithm
C04
Output during warm-up 0: Held at 4 mA
up
1: Held at 20 mA
2: Set value remains held.
Held at 4 mA.
C05
Output during
0: Not held
maintenance
1: Held output just before
Held output just before
maintenance service.
maintenance service.
2: Set value remains held.
C06
Output during
0: Not held
Held output just before
calibration 1: Held output just before
calibration.
calibration.
2: Set value remains held.
C07
Output in abnormal 0: Not held
Held output at a preset
state
1: Held output just before
value.
abnormal state occurs.
2: Set value remains held.
C11
Min. oxygen
See Section 8.1.
% O
2
0% O
2
concentration
C12
Max. oxygen
See Section 8.1.
% O
2
25% O
2
concentration
C13
Minimum humidity
See Section 8.1.
% H
2
O 0% H
2
O
C14
Maximum humidity
See Section 8.1.
% H
2
O 25% H
2
O
C15
Minimum mixing ratio See Section 8.1. kg/kg
0 kg/kg
C16
Maximum mixing ratio See Section 8.1. kg/kg
0.2kg/kg
C30
Output damping
0 to 255
Seconds
0 second
factor
C31
Set value during
2.4 to 21.6
mA
4 mA
warm-up
C32
Set value during
2.4 to 21.6
mA
4 mA
maintenance
C33
Set value during
2.4 to 21.6
mA
4 mA
calibration
C34
Set value in abnormal 2.4 to 21.6
mA
3.4 mA
state
T10.6C.EPS