Caution – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 125

IM 11M12A01-05E
9-8
9.2.1.6Setting
When setting calibration timing requirements, bear the following precautions in mind:
CAUTION
(1) If the calibration interval is shorter than the sum of stabilization time plus calibration
time, the second calibration start time will conflict with the first calibration. In such a
case, the second calibration will not be conducted. (When both zero and span
calibrations are to be performed, the calibration time is double that required for a
single (zero or span) calibration.)
(2) For the same reason, if the calibration start time conflicts with manual calibration or
semi-automatic calibration, the current calibration will not be conducted.
(3) If the calibration time conflicts with maintenance service or blowback operations,
calibration will start after completing the maintenance service or blowback opera-
tions (see Section 8.2.1, earlier in this manual).
(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration
will be conducted; a second or later calibration will not be conducted.
(5) If a past date is set to the calibration start day, no calibration will be conducted.
Table 9.1 Parameter Codes for Calibration Setting
Set Item
Parameter code Set value
Engineering unit
Zero-gas concentration
B01
Set Zero-gas concentration
%O
2
Span-gas concentration
B02
Set Span-gas concentration
%O
2
Calibration mode
B03
0 : Manual calibration
1 : Semi-automatic and manual
2 : Automatic, semi-automatic, and manual
Output stabilization
B04
0 minutes 0 seconds to 60 minutes 59 seconds
MM.SS
Calibration time
B05
0 minutes 0 seconds to 60 minutes 59 seconds
MM.SS
Calibration period
B06
0 days 0 hours to 255 days 23 hours
Date and time
Start date and time
B07
Date and time of first calibration
YY.MM.DD.HH.MM
Calibration procedure
B08
0 : Zero and span
1 : Span only
2 : Zero only
T9.1.EPS
9.2.1.7Default Values
When the analyzer is delivered, or if data are initialized, the calibration settings are by
default, as shown in Table 9.2.
Table 9.2 Default Settings for Calibration
Item
Default Setting
Calibration mode
Manual
Calibration procedure
Span - zero
Zero-gas (oxygen) concentration
1.00%
Span-gas (oxygen) concentration
21.00%
Output hold (stabilization) time
10 minutes, 00 seconds
Calibration time
10 minutes, 00 seconds
Calibration interval
30 days, 00 hours
Start day and time
00 (YY) 01 (MM) 01(DD) 00:00
T9.2.EPS