B1 0, Span y, Open/20.84 – Yokogawa Integral Oxygen Analyzer ZR202 User Manual
Page 104: Zero y, Open/0.89, Calend
IM 11M12A01-05E
8-7
8. Detailed Data Setting
F8.1E.EPS
Switch operations
Display
Output hold time
during calibration
>
ENT
b
1
0
>
ENT
CAL
>
ENT
CAL
>
ENT
SPAn
Y
>
ENT
21.00
%
>
ENT
OPEn/20.84
>
ENT
20.84
%
>
ENT
ZEro
Y
>
ENT
0.98
%
>
ENT
OPEn/0.89
>
ENT
0.89
%
>
ENT
CALEnd
>
ENT
b
10
>
ENT
Measured-value display
>
>
>
>
>
>
>
>
>
>
>
>
>
>
Figure 8.1 Definition of during calibration
During semi-automatic calibration, “under calibration” is the time, starting when a
calibration instruction is executed with an infrared switch or a contact input, to make a
series of calibrations, until the preset output stabilization time elapses.
During automatic calibration, “under calibration” is the time, starting when automatic
calibration is carried out at the calibration start time, until the preset output stabilization
time elapses.
(4) “Error” appears when Error 1 to Error 4 are being issued