IM 11M12A01-05E
10-13
10. Other Functions
Calibration-related Items in Group B
Code
Item
Tuning
Engineering unit Default setting
B01
Zero-gas concentration
0.3 to 100
% O
2
1% O
2
B02
Span-gas concentration
4.5 to 100
% O
2
21% O
2
B03
Calibration mode
0: Manual calibration
Manual calibration
1: Semi-automatic and manual
calibration
2: Automatic, semi-automatic, and
manual calibration
B04
Output stabilization time
0 minutes, 0 seconds to 60 minutes, 59
MM.SS
10 minutes, 0 seconds
seconds
B05
Calibration time
0 minutes, 0 seconds to 60 minutes, 59
MM.SS
10 minutes, 0 seconds
seconds
B06
Calibration period
0 days 0 hours to 255 days 23 hours
DD.HH
30 days, 0 hours
B07
Start time
YY.MM.DD/HH.MM 00.01.01.00.00
B08
Calibration
0: Zero and span
Zero and span
1: Span only
2: Zero only
B09
Calibration concentration
Display only
% O
2
measurement
B10
Manual calibration
B11
Semi-automatic calibration
T10.6B.EPS