beautypg.com

1 power wiring, 2 grounding wiring, 5 power and grounding wiring – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 90

background image

IM 11M13A01-02E

5-11

5. Wiring

5.5

Power and Grounding Wiring

This wiring supplies power to the converter and grounds the converter/detector.

Converter

Jumper plate

L N G

Detector

Ground

100 - 240VAC

50/60Hz

Grounding to the ground terminal

on the converter case

Converter case

FG terminal

Lock washer

Crimp-on terminal of

the ground wire

F5.9E.EPS

ZR402G

ZR22S

Figure 5.9 Power and Grounding wiring

5.5.1

Power Wiring

Connect the power wiring to the L and N terminals of the converter. Proceed as follows:

(1) Use a 2-core or a 3-core shielded cable.
(2) The size of converter terminal screw threads is M4. Each cable should be terminated

corresponding crimp-on terminals.

5.5.2

Grounding Wiring

The ground wiring of the detector should be connected to the ground terminal of the
detector case. The ground wiring of the converter should be connected to either the
ground terminal of the converter case or the protective ground terminal in the equip-
ment. The grounding terminals of the detector are of size of M4 (inside) and M5
(outside), and the grounding terminals of the converter are of size M4. Proceed as
follows:

(1) Keep ground resistance to 100

⍀ or less.

(2) When the ambient temperature of the wiring installation is 80 to 150

Њ C for the wiring

of the detector, use wiring material with sufficient heat resistance.

(3) When connecting the ground wiring to the ground terminal of the converter case, be

sure that the lock washer is in contact with the case surface (see Figure 5.9).

(4) Ensure that the jumper plate is connected between the G terminal and the protective

ground terminal of the converter.