3 check valve (parts no. k9292dn or k9292ds) – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 47

IM 11M13A01-02E
2-28
2.7.3
Check Valve (Part No. K9292DN or K9292DS)
This valve is mounted on the calibration gas line (directly connected to the detector).
This is applied to a system based on the (System 2 and 3) system configuration .
This valve prevents the process gas from entering the calibration gas line. Although it
functions as a stop valve, operation is easier than a stop valve as it does not require
opening/closing at each calibration.
Screw a check valve, instead of a stop valve into the calibration gas inlet of the detector.
Standard Specifications
Connection:
Rc1/4 or 1/4NPT
Material:
SUS304 (JIS)
Pressure:
70 kPa G or more and 350 kPa G or less
Weight:
Approx. 40 g
Joint: Rc 1/4, Material: SUS304 (JIS)
Joint: 1/4 NPT, Material: SUS304 (JIS)
T2.10E.EPS
Part No.
K9292DN
K9292DS
Description
K9292DN : Rc1/4 (A part), R1/4 (B part)
K9292DS : 1/4NPT female (A part),1/4NPT male (B part)
F2.11E.EPS
Approx.19
Approx.54
A
B
Unit : mm