2 zo21p-h adapter for high-temperature probe – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 29

IM 11M13A01-02E
2-10
2.3.2
ZO21P-H Adapter for High-Temperature Probe
The probe adapter is used to lower the sample gas to a temperature below 700
ЊC (below
300
ЊC at probe adapter surface) before it is fed to the detector.
Insertion length:
1 m, 1.5 m
Material in Contact with Gas:
SUS 316 (JIS), Zirconia, SiC or SUS 310S, SUS 304
(JIS) (flange)
Probe Material:
SiC, SUS 310S (JIS)
Installation:
Flange mounting (FF type or RF type)
Probe Mounting Angle:Vertically downward within
؎5؇
Where the probe material is SUS 310S, horizontal mounting is
available.
Construction:
Non explosion-proof. Rainproof construction
Weight:
Insertion length of 1.0 m:
approx. 6.5 kg (JIS) / approx. 8.5 kg
(ANSI)
Insertion length of 1.5 m:
approx. 7.5 kg (JIS) / approx. 9.5 kg
(ANSI)
Sample gas temperature: 0 to 1400
؇ C (when using SiC probe) 0 to 800؇ C (when using
SUS 310S probe adapter)
Sample gas pressure:
-0.5 to + 5 kPa. When using in the range of 0 to 25 vol% O
2
or
more, the sample gas pressure should be in the range of -0.5 to +0.5
kPa. (Where the sample gas pressure for the high-temperature probe
is negative, an auxiliary ejector is necessary.)
-100
-150
*A
-J
-N
-M
-L
-A
-R
-Q
-T
-S
-E
-H
-A
-B
T2.3E.EPS
Model
Suffix code
Option
code
ZO21P
Material
Insertion
Length
Flange
High Temperature Probe Adapter
SiC
SUS 310S
1.0 m
1.5 m
JIS 5K 50 FF SUS304
JIS 10K 65 FF SUS304
JIS 10K 80 FF SUS304
JIS 10K 100 FF SUS304
ANSI Class 150 4 RF SUS304
ANSI Class 150 2 1/2 RF SUS304
ANSI Class 150 3 RF SUS304
JPI Class 150 3 RF SUS304
JPI Class 150 4 RF SUS304
DIN PN10 DN50 A SUS304
Style A
Description
Style code
Note: For this high-temperature use probe adapter, be sure to
specify the ZR22S probe of its insertion length 0.15 meters.