3-18

3.6

Insulation Resistance Test

Even if the testing voltage is not so great that it causes dielectric breakdown, testing
may cause deterioration in insulation and a possible safety hazard. Therefore, conduct
this test only when it is necessary.

The applied voltage for this test shall be 500 V DC or less. The voltage shall be applied
for as short a time as practicable to confirm that insulation resistance is 20 M

⍀ or more.

Remove wiring from the converter and the detector.

1. Remove the jumper plate located between terminal G and the protective grounding

terminal.

2. Connect crossover wiring between L and N.
3. Connect an insulation resistance tester (with its power OFF). Connect (+) terminal to

the crossover wiring, and (-) terminal to ground.

4. Turn the insulation resistance tester ON and measure the insulation resistance.
5. After testing, remove the tester and connect a 100 k

⍀ resistance between the cross-

over wiring and ground, to discharge.

6. Testing between the heater terminal and ground, contact output terminal and ground,

analog output/input terminal and ground can be conducted in the same manner.

7. Although contact input terminals are isolated, insulation resistance test cannot be

conducted because the breakdown voltage of the surge-preventing arrester between
the terminal and ground is low.

8. After conducting all the tests, replace the jumper plate as it was.

1

FG

2

AO1

(+)

3

AO1

(-)

4

AO2

(+)

5

AO2

(-)

6

CJ
(+)

7

CJ

(-)

8

TC

(+)

9

TC

(-)

10

CELL

(+)

11

CELL

(-)

12

FG

13

DI-1

14

DI-2

15

DI-C

16

AI

(+)

17

AI
(-)

18

AC-

ZERO

19

AC-

SPAN

20

AC-

COM

21

FG

22

FG

CELL

(+)

CELL

(-)

TC
(+)

TC

(-)

CJ
(+)

CJ

(-)

23

DO-1

24

DO-1

25

DO-2

26

DO-2

27

DO-3

28

DO-3

29

DO-4

30

DO-4

31

HTR

32

HTR

33

L

34

N

35

G

36

HTR

HTR

Contact

input 2

Contact

input 1

Solenoid valve

for automatic

calibration

Solenoid valve for

automatic calibration

Span gas

Zero gas

Automatic Calibration Unit

Separate-type Zirconia Oxygen Analyzer, ZR402G Converter

Separate-type Zirconia Oxygen Analyzer,
ZR22S Detector

Insulation

resistance tester

Insulation

resistance tester

+

-

+

-

Insulation

resistance tester

Remove
jumper plate

Remove wiring to connect external circuits
and Connect crossover wiring

F3.19E.EPS

+

-

crossover wiring

crossover wiring

Figure 3.19 Insulation Resistance Test