beautypg.com

Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual

Page 78

background image

IM 11M13A01-02E

4-9

4. Piping

The following parts are required for blowback piping.

• Blow pipe (to be prepared as illustrated in Figure 4.9.)
• Two-way solenoid valve: “ Open “ when electric current is on. (Found on the open

market)

• Air set (recommended by YOKOGAWA, K9473XH / K9473XJ or G7004XF /

K9473XG)


Manufacture the blow pipe as illustrated in Figure 4.9, and mount it on the high-
temperature probe adapter.

Approximately
200

Rc1/4

Blind flange of the adapter for
high temperature probe

Welded

8 (O.D.) by 6 (I.D.) Stainless steel pipe

F4.9E.EPS

Unit : mm

Figure 4.9 Manufacturing Blow pipe