1 general specifications – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 21

IM 11M13A01-02E
2-2
2.1
General Specifications
Standard Specifications
Measured Object:
Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inflammable gases. May not be applicable corrosive gas
such as ammonia and chlorine.)
The sampling gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer. In this case,
contact with YOKOGAWA and its agency.
Measurement System: Zirconia system
Oxygen concentration: 0.01 to 100 vol% O
2
Output Signal:
4 to 20 mA DC (maximum load resistance 550
⍀)
Measurement Range:
Any setting in the range of 0 to 5 through 0 to 100 vol% O
2
(in
1 vol% O
2
), or partial range
Digital Communication (HART): 250 to 550
⍀, depending on number of field devices
connected to the loop (multi-drop mode).
Note:
HART is a registered trademark of the HART Communication Foundation.
Display Range:
0 to 100 vol% O
2
Warm-up Time:
Approx. 20 min.
Repeatability:
؎0.5% Maximum value of range setting. (range up to 0 to 25 vol%
O
2
)
؎1% Maximum value of range setting. (range from 0 to 25 vol%
O
2
up to 0 to 100 vol% O
2
)
Linearity:
(Excluding standard gas tolerance)
(Use oxygen of known concentration (in the measuring range) as
the zero and span calibration gas.)
؎1% Maximum value of range setting ; from 0 to 5 vol% O
2
to 0
to 25 vol% O
2
range
(Sample gas pressure: within
؎4.9 kPa)
؎3% Maximum value of range setting ; from 0 to 25 vol%O
2
to 0
to 50 vol% O
2
range
(Sample gas pressure: within
؎0.49 kPa)
؎5% Maximum value of range setting ; from 0 to 50 vol% O
2
to 0
to 100 vol% O
2
range (Sample gas pressure: within
؎0.49 kPa)
Drift:
(Excluding the first two weeks in use)
Both zero and span
؎2% Maximum value of range setting/month
Response Time :
Response of 90% within 5 seconds. (Measured from when gas is
introduced from calibration gas inlet and analog output start
changing.)