11.2.3

Adjust LCD screen contrast

An LCD is built in the ZR402G converter. The contrast of this LCD is affected by its
ambient temperature. For this reason, the LCD is shipped, after adjusting the contrast so
as to become the most suitable in a room temperature(20-30

؇ C). However, when display

on the LCD is hard to see, adjust the LCD contrast by change the resistance of the
variable resistor; its position is shown in Fig. 11.8

Variable resistor

Figure 11.8