4 default values, Alarms setup, Oxygen alarms – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 126

IM 11M13A01-02E
8-9
8. Detailed Data Setting
Parameter: Oxygen
᭜
Hysteresis: 0.1%O2
᭜
Contact delay: 3 s
᭜
Setpoints
Enter
Alarms setup
F8.5E.EPS
High High alarm: OFF
᭜
Set value: 1 0 0 . 0 % O2
᭜
High alarm: ON
᭜
Set value: 1 0 0 . 0 % O2
᭜
Low alarm: OFF
᭜
Set value: 0 . 0 % O2
᭜
Low Low alarm: OFF
᭜
Set value: 0 . 0 % O2
Enter
Oxygen alarms
F8.6E.EPS
Figure 8.5 Alarms Setup Display Figure 8.6 Oxygen Alarms Display
8.3.4
Default Values
When the analyzer is delivered, or if data are initialized, the default alarm set values are
as shown in Table 8.3.
Table 8.3 Alarm Setting Default Values
Set item
Set value
Hysteresis
0.1% O
2
Delay time
3 seconds
High-high limit alarm
Off
High-high-limit alarm setpoint
100% O
2
High-limit alarm
Off
High- and low-limit alarm setpoints
100% O
2
Low-limit alarm
Off
Low-limit alarm setpoint
0% O
2
Low-low-limit alarm
Off
Low-low-limit alarm setpoint
0% O
2
T8.3E.EPS