2 setting procedure, 3 default values – Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 132

IM 11M13A01-02E
8-15
8. Detailed Data Setting
8.5.2
Setting Procedure
The following are set so that semi-automatic calibration starts when input contact open
is applied to “Input1.”
Proceed as follows:
(1) Press the Setup key in the Basic panel display to display the Execution/Setup display.
(2) Select Setup in the Execution/Setup display to display the “Commissioning” (Setup)
display.
(3) Select “Contact setup” in the “Commissioning” (Setup) display.
(4) Select “Input contacts” in the Contact setup display. The Input contacts display then
appears, as shown in Figure 8.11.
(5) Select Input1 in that display. The Input contacts display then appears, as shown in
Figure 8.12.
(6) Select Calibration start.
(7) Select Input1 “closed.” An Open or Closed selection display then appears.
(8) Choose Open.
Input1: No function
᭜
Input1 contact is active at:
Closed
᭜
Input2: No function
᭜
Input2 contact is active at:
Closed
Enter
Input contacts
F8.11E.EPS
Input1: No function
᭜
Input1 contact is active at:
Closed
᭜
Input2: No function
᭜
Input2 contact is active at:
Closed
Enter
Input contacts
No function
Cal. gas press. low
Range change
Calibration start
Process up set
Blow back start
F8.12.EPS
Figure 8.11 Input Contacts Display Figure 8.12 Input Contacts Display
8.5.3
Default Values
When the analyzer is delivered, or if data are initialized, contact input is disable.