Replacement of heater strut assembly
Refer to Figure 11.4 as an aid in the following discussion.
Remove the cell assembly
6
, following Section 11.1.2, earlier in this manual. Open the
terminal box
14
and remove the three terminal connections – CELL +, TC + and TC -.
Before disconnect the HTR terminals, remove the terminal block screw
26
. Keeping the
other terminal remaining to be connected. Disconnect the two HTR connections. (These
terminals have no polarity.)
Remove the four bolts
10
and terminal box
14
with care so that the already discon-
nected wire will not get caught in the terminal box.
Loosen Screw
17
until Heater Strut Assembly
21
plate can be removed.
There’s no need to remove O-ring
16
which prevents Screw
17
from coming out.
Pull out connector
12
.
Loosen and remove the screw
8
with a special wrench (part no. K9470BX or equiva-
lent) and then remove the heater strut assembly
21
from the detector
22
.
To reassemble the heater strut assembly, reverse the above procedure:
Insert the heater strut assembly
21
into the detector
22
, while inserting the calibration
pipe in the detector
22
into the heater section in the heater strut assembly
21
as well as
in the bracket hole. Coat the screw (8) with grease (NEVER-SEEZ: G7067ZA) and
tighten the screw
8
with a special tool (part no. K9470BX or equivalent) with a
tightening torque of 12N • m
±
10 percent.
Next, to install the O-rings
20
on the calibration-gas and reference-gas pipes, disas-
semble the connector
12
in the following procedure:
First, remove the screw
23
and then remove the plate
15
and two caps
18
. If the O-ring
20
remains in the hole, pull them out from the back. Pass the heater and thermocouple
leadwire through the connector
12
. Also, pass the calibration-gas and reference-gas
pipes through the opening of the connector
12
. If the O-ring
20
fails, replace it with a
new one.
Push the two caps
18
into the associated opening of the connector
13
. Insert the plate
(15), aligning it with the groove of the cap
18
, and tighten it with the screw
23
. If you
attempt to insert the calibration-gas and reference-gas pipes into the connector
12
without disassembling the connector
12
, the O-ring may be damaged. Tighten the
Tighten Screw
17
in Heater Strut Assembly
21
until connector
12
can’t move.
Reassemble in reverse order to the above disassembly procedure.
The two wires with ceramic insulators from the heater strut assembly are heater wires,
and the single-core shielded wire is the cell signal + terminal; for the two-core shielded
cable, the semi-translucent rubber-sheathed wire is the thermocouple + terminal, and the
other wire is the
Ϫterminal. (If the wires are marked, match the markings with those on
the terminal board).
When installing the cell assembly
6
, replace the metal O-ring
7
with a new one.