Yokogawa Single Channel Oxygen Analyzer System ZR22/ZR402 User Manual
Page 163

IM 11M13A01-02E
10-12
Table 10.5 Initialization Items and Default Values
T10.5E-1.EPS
Item
Data to be initialized
Default setting
Type of equipment
Not initialized
Equipment
Detector
ZR22
selection
Measurement gas
Wet gas
Display item
1st display item
Oxygen concentration
2nd display item
Current output 1
3rd display item
Current output 2
Tag name
Deleted
Trend graph
Parameter
Oxygen concentration
Sampling interval
30 seconds
Displayed data
Upper limit (graph)
25%O
2
Lower limit (graph)
0%O
2
Automatic return time
0 min.
Language
Not initialized
Mode
Manual
Calibration procedure Span - zero
Zero-gas concentration 1.00%O
2
Span-gas concentration 21.00%O
2
Calibration data
Calibration setting
Output hold time
10 min., 00 sec.
Calibration time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
00:00
Mode
No function (invalid)
Blowback
Blowback setting
(Output) hold time
10 min., 00 sec.
Blowback time
10 min., 00 sec.
Interval
30 days, 00 hr.
Start date
01/01/00
Start time
00:00
Parameter
Oxygen concentration
mA-output 1
Min. oxygen concentration 0%O
2
mA-output 2
Max. oxygen concentration 25%O
2
@ @
Output damping
0%
Output mode
Linear
Warm-up
4 mA
Set value
4 mA
Current output
Maintenance
Previous value held
data
Output hold setting
Set value
4 mA
Calibration, blowback
Previous value held
Set value
4 mA
Error
Preset value held
Set value
4 mA