Performance verification test record – Teledyne LeCroy WaveLink Differential Probe Series (13-25 GHz) User Manual
Page 76

WaveLink Series Differential Probe (13-25 GHz)
70
WL-HBW-A-OM-E Rev B
NOTE: The fast edge doesn't have a very flat response, and the deskew fixture is not designed for
such high bandwidth, so the aberrations seen here are mostly determined by the source and not the
probe.
Performance Verification Test Record
This record can be used to record the results of measurements made during the performance
verification of the WaveLink Series of Differential Probes. Photocopy this page and record the results
on the copy. File the completed record as required by applicable internal quality procedures. The
section in the test record corresponds to the parameters tested in the performance verification
procedure. The numbers preceding the individual data records correspond to the steps in the
procedure requiring the recording of data.
Results to be recorded in the column labeled Test Result are the actual specification limit check. The
test limits are included in all of these steps. Other measurements and the results of intermediate
calculations that support the limit check are to be recorded in the column labeled Intermediate
Results.
Permission is granted to reproduce these pages for the purpose of recording test results.
NOTE: Use a new Test Record for each tested probe, probe tip module, and lead.