Teledyne LeCroy WaveLink Differential Probe Series (13-25 GHz) User Manual

Page 62

background image

WaveLink Series Differential Probe (13-25 GHz)

56

WL-HBW-A-OM-E Rev B

8.

At this point, the Calibrator must be set up before performing the functional tests. Select

Utilities → Utilities Setup...

from the menu bar.

9.

Touch the

Aux Output

tab as follows:

10.

Touch the

Square

button to obtain a square wave output signal.

Auxiliary output setup

NOTE: TTL is processed internally (and the Aux Output dialog shows a TTL checkbox) for the
following oscilloscope models: WaveMaster 5000, 6000, SDA 4000, 5000, and 6000.

11.

Set the amplitude to 1 Volt, Frequency to 1.00 kHz.

12.

Verify the screen indicates the proper settings in the Aux Output fields.

13.

Set the probe sensitivity to 200 mV/div.

Connect the + tip of the interconnect lead module to the center connector of the calibrator
output signal, and the lead to the shell (ground) of the connector. If necessary, readjust the
tip spacing.