beautypg.com

Teledyne LeCroy WaveLink Differential Probe Series (13-25 GHz) User Manual

Page 69

background image

Operator's Manual

WL-HBW-A-OM-E Rev B

63

Output Zero Voltage

9.

Initiate an AutoZero.

10.

Wait an additional 15 minutes, then record the DMM reading to 1 mV resolution on the

Performance Verification Test Record (on

page 70

)

as

Output Voltage after AutoZero

(step 2).

11.

Take the difference of the two readings recorded steps 8 and 10 (previous) and multiply by
the Effective gain value shown in the channel attributes when the probe is attached. See

Operation with a Teledyne LeCroy Oscilloscope (on

page 19

)

for more information.

12.

On the

Performance Verification Test Record (on

page 70

)

, record the result as

Output

Zero

(step 3).

13. Verify the absolute value of Output Zero is less than 10 mV.