Verify the screen shows a square wave (you may need to adjust the channel offset). Refer to

the following screen-shot. If no square wave is shown, the

+

channel of the probe may be

faulty.

Square wave output signal

16.

Make desired adjustments to your oscilloscope settings and obtain a stable display.

17.

Adjust the tips so both touch or are almost touching.

18.

Connect both tips to the calibrator output signal.

19.

Verify a straight line is shown centered on the screen. There should be no vertical deflection,
to indicate good CMRR.

20.

One of the channels may be at fault if a square wave or part of a square wave is shown.

This concludes the functional tests of the WaveLink High Bandwidth Differential Probe.