The Utility Program screen is then shown. Use the steps shown on the screen as follows to set your
Pogo-Pin Tip Ω values correctly:

1.

Disconnect your Dxx05-PT Positioner Tip Browser (and any other probes) from your
oscilloscope and Exit the oscilloscope application using

File → Exit

from the menu bar. The

desktop of your oscilloscope should be showing with the instrument still running.

2.

Now, start the oscilloscope application by touching the desktop shortcut.

3.

Connect the Platform/Cable Assembly of your Positioner Tip Browser to the upper-row
Channel 1 oscilloscope input using the LPA-2.92 mm to ProLink Probe Adapter.

Confirm your Probe dialog is shown as part of the Channel 1 dialog.

4.

Provide the PT Tip Resistance value. The label on the vial containing your replacement pins
shows this resistance value.

5.

Click the Program Probe button to proceed with your tip reprogramming. When the Status
message shows DONE!, the reprogramming procedure has finished.