2. INTRODUCTION 

2.1 9060Z TRANSMITTER 

The Teledyne Analytical Instruments 9060Z oxygen transmitter is designed to measure the oxygen gas 
concentration  in  combustion  applications  such  as  power  stations  and  industrial  boilers.  The 
measurement  is  used  to  control  the  air  /  fuel  ration  in  the  burner  to  maintain  efficiency  and  safe 
operation conditions. 
The transmitter is designed to be used in conjunction with a Teledyne Analytical Instruments probe 
model  9060H  or  9060UL/UH.  These  two  probes  use  a  zirconia  sensor  that  must  operate  at  >650°C. 
The 9060H probe has an integrated heater, and the 9060UL/UH probe uses the process heat to heat 
the  sensor.  There  is  a  unique  cable  connection  between  the  9060Z  transmitter  and  the 
9060H/9060UL/UH probe. 
The model 9060Z oxygen transmitter is based on the well known model 9060H transmitter. It includes 
a number of hardware and software improvements such as a graphic display, larger characters, faster 
microprocessor,  simplified  set  up  menu,  alarm  logging,  faster  probe  heater  control  and  more 
calculated values such as burner efficiency. 
The  9060Z  Oxygen  Transmitter  has  a  variety  of  user‐selectable  functions.  They  are  simple  to  use 
because each selection is menu driven. For options you are not sure about, read the manual on that 
particular item in chapter 6, Setup Mode. 

Features include:‐ 
Inputs 
Two zirconia oxygen probes, heated or unheated 
Oxygen 

read 0‐1% to 0‐100%

 

Furnace, kiln or flue thermocouple, field selectable as type K, J, R or S 
Main flame established safety interlock (for heated probes only) 
Purge pressure or flow switch 

Outputs 
Two linearised 4‐20mA or 0‐20mA DC isolated outputs, max. load 1000Ω 
The output function and the range are field selectable 
Common alarm relay 
Three other alarm relays with selectable functions 

Digital Interface 
RS‐232 or two wire RS‐485 MODBUS  for connection to a computer/DCS/PCL for diagnostics of the 
transmitter, probe or combustion process. 

Display 
Multi font graphical display 
Large font characters for the oxygen on the top line 
Selectable lower line items for the secondary display functions. ie Probe temperature, Oxygen second 
probe Alarm display mode that shows the time of the alarm, the acceptance time and the time that 
the alarm was cleared of up to 4000 alarm events 

2.2 SERIES 9060H OXYGEN PROBES & SENSORS 

Please note that if this equipment is not installed and used in the manner described in this manual then 
the safety protection provided by the equipment may be impaired.