beautypg.com

Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 6

background image

Model 9060Z Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Important Notice Regarding 9060 Probe Option ‐ FIL‐3 

 

WARNING: The only identifiable standard for flame arresters for general use is British Standard BS 
EN 12874:2001. British Standard BS EN 12874:2001 refers to an operating environment up to 150 
Degrees Centigrade. 

The  FIL‐3  device  optionally  fitted  to  9060  Heated  Zirconia  Probes  (the  “Probes"  or  "Probe") 
operate in an environment considerably greater than 150 Degrees Centigrade. 

Therefore,  we  know  of  no  Australian,  British,  European  or  USA  standard  applicable  to  flame 
arresters or their testing above 150 degrees Centigrade. Consequently, the FIL‐3 device cannot be 
certified as a safety device. 

The probe is only on probes during the start up processes of a combustion appliance. 

The Teledyne Analytical Instruments Burner Interlock Relay facility, which is a standard part of the 
Teledyne  Analytical  Instruments  transmitter,  is  designed  to  be  wired to  the  main  safety  shut‐off 
fuel valves in a way that can shutdown the probe heater when the fuel valves are closed. 

The risk of ignition of flammable gas mixture at the hot end of the Probe can only be minimized by 
correct  use,  maintenance  and  operation  of  the  FIL‐3  device.  The  user  of  the  FIL‐3  device  is 
responsible for verification and maintenance and correct use and operation of the FIL‐3 device. 

THE USER AGREES THAT IT USES THE PROBE AND THE FIL‐3 DEVICE AT ITS SOLE RISK. TELEDYNE 
ANALYTICAL INSTRUMENTS TO THE FULL EXTENT PERMITTED BY LAW, GIVES NO WARRANTIES OR 
ASSURANCES AND EXCLUDES ALL LIABILITY (INCLUDING LIABILITY FOR NEGLIGENCE) IN RELATION 
TO THE PROBE AND THE FIL‐3 DEVICE. 

The  user  must  ensure  that  it  correctly  follows  all  instructions  in  relation  to  the  Probe  and  FIL‐3 
device, correctly understands the specifications of the Probe and FIL‐3 device and ensures that the 
Probe and FIL‐3 device are regularly inspected and maintained. 

FIL‐3 equipped Probes should be inspected at least once a year for corrosion and more frequently 
if there is any reason to suspect that corrosion may have occurred.