8. CALIBRATION MODE 

This  chapter  describes  the  functions  available  when  the  CALIBRATION  mode  is  selected  on  the 
transmitter. 
For specific information about calibrating the transmitter see chapter 10, Instrument Calibration. 
The  CALIBRATION  mode  is  accessed  by  pressing  and  holding  the  SETUP  key  for  approximately  4 
seconds until the words “Calibration Menu” appears at the bottom of the display. The transmitter will 
return to the RUN mode when the SETUP key is pressed again or 60 seconds after the last key press. 
Changing options in the CALIBRATION mode is the same as the SETUP mode. See chapter 6.2 Changing 
Options. 

8.1 FUNCTION SUMMARY TABLE 

When the transmitter is in the CALIBRATION mode the SETUP light will be on and the words “Commissioning 
Menu” will be at the bottom of the display. The following table shows the 

CALIBRATION menu functions: 

Menu 

#  

Function name  

Range 

Default 

value 

Reference Voltage 1, 50mV  

40.00 to 60.00mV  

47.14mV 

Reference Voltage 2, 200mV  

180.00 to 210.00mV  

182.24mV 

Reference Voltage 3, 1200mV  

1150.0 to 1250.0mV  

1221.8mV 

Reference Voltage 4, 2500mV  

2400.0 to 2550.0mV 

2489.2mV 

Output Channel 1, Calibration  

Auto Calibrated Manual Calibrated Set 
4mA Set 20mA 

Auto 
Calibrated 

Output Channel 1, 4mA Trim  

3.00 to 5.00mA  

4.00mA 

Output Channel 1, 20mA Trim  

19.00 to 21.00mA  

20.00mA 

Output Channel 2, Calibration  

Auto Calibrated Manual Calibrated Set 
4mA Set 20mA  

Auto 
Calibrated 

Output Channel 2, 4mA Trim  

3.00 to 5.00mA 

4.00mA 

10 

Output Channel 2, 20mA Trim  

19.00 to 21.00mA 

20.00mA 

11 

Ambient Temperature Sensor Offset  ‐10.0°C to 10.0°C  

 0.0°C 

12 

Low Oxygen Cal 1  

80.0% to 120.0% 

100.00% 

13 

Low Oxygen Cal 2  

80.0% to 120.0%  

100.00% 

14 

Transmitter Output Select  

4‐20mA / 0‐20mA / 4‐20mA over‐
scale limit / 0‐20mA over scale limit 

 4‐20mA 

15 

Transmitter Output Limiting  

Disabled / Hold 0mA / Hold 4mA / 
Hold 20mA  

Hold 0mA 

16 

Mains Voltage Detection Override  

Automatic / 220/240 / 10/120  

Automatic 

17 

Mains Frequency Detection Override Automatic / 50Hz / 60Hz  

Automatic 

18 

Heater SSR Selection Normal  

Heater1 <‐> Heater2 Heaters <‐> Cal 
Purge 

Normal 

19 

SSR Fail Protection  

Enabled / Disabled  

Enabled