6.3 SETUP MODE FUNCTIONS 

 

6.3.1 Probe 1 Offset 

Each  Teledyne  Analytical  Instruments  probe  has  an  offset  calibration  value  printed  on  a  tag  that  is 
attached to the probe when it is dispatched. To achieve accurate measurement the offset value must 
be  entered  with  the  same  polarity  as  it  is  printed  on  this  label.  The  offset  value  should  be  within 
±1.0mV 

NOTE:  An  offset  of  1.0mV  will  change  the  oxygen  reading  by  approximately  1%  oxygen  when  the 
probe is in ambient air. However, as the process oxygen measurement drops, this offset will have a 
reduced affect. At a process gas oxygen concentration of 2%, the 1.0mV offset error will only change 
the reading by 0.1% oxygen. 
If in any doubt about the correct offset value, set it to 0.0mV. 

The function ’Probe 2 Offset’ will only appear if the transmitter has been configured for 2 probes. (see 
chapter 7.2.3 Number of Probes ) 
 

6.3.2 Lower Line Items 

This function allows the operator to change the items that are available to be displayed on the lower 
line of the transmitter in RUN mode. If the word “Enabled” appears on the display for a selected lower 
line  option,  the  measurement  will  be  available  to  be  shown  on  the  display  in  the  RUN  mode  by 
scrolling through the list using the DISPLAY UP and DISPLAY DOWN keys. 
A lower line measurement can be “Enabled” or disabled by pressing the ENTER key.