14. MAINTENANCE 

14.1 TRANSMITTER MAINTENANCE 

The 9060Z transmitter has several hardware checking systems that confirm the correct operation of 
the transmitter and raise an alarm if there is a fault detected. Because of this system, the transmitter 
only requires periodic maintenance. Teledyne Analytical Instruments recommends an annual check of 
the transmitter. 

The annual check should include the following items – 

1. Measure  and  record  the  reference  voltages,  and  check  that  the  correct  voltages  are  set  in  the 

calibration menu 

2. Check that the reference air flow from the port on the bottom of the transmitter case is between 

100 and 300cc/m 

3. Check for contamination and potential blockage of the reference air filter that is mounted on the 

bottom of the transmitter case 

4. If an external burner interlock has been connected (terminals 10 & 11), check that if one of the 

wires is disconnected the probe heater is disabled (9060H probes only) 

5. Check  the  integrity  of  the  cable.  Particularly check  the main  power  supply  cable  and  the  probe 

cable for any damage. 

6. Check the integrity of the earth connections in the transmitter 

14.2 CLEANING  

The 9060Z transmitter is made in an IP65 / IP54 rated diecast case. It is painted with a tough ripple 
finish power coating to maintain a clean look. However, if it is necessary to clean the case use either a 
warm soapy water solution or an isopropyl alcohol. 
A build up of dust on or around the transmitter should be removed before the cabinet door is opened. 
If there is a buildup of dust inside the cabinet check the door o‐ring seal for damage and replace it if 
necessary. 

 

Take care when wiping the display window to avoid scratching the surface. 
If  the  filters  of  the  probe  (if  fitted)  become  blocked,  use  an  ultrasonic  cleaner  with  a  few  drops  of 
household detergent. If the filters cannot be cleaned this way replace the sintered filters. 
The  filter  in  the  base  of  the  analyzer  may  become  blocked  if  the  analyzer  is  operating  in  a  dusty 
environment. It is available as a field replacement item. 

14.3 REPLACEMENT PARTS 

The following list identifies field replaceable parts – 

Stock number  

Description 

18020 

1732‐1 Main PCB  

18021 

1732‐2 Display PCB  

15039 

MP‐24E Reference air pump  

14105 

Fuse ‐M205 3.15A 250V fuse (pack of 5)  

14093 

Instruction manual  

18064 

EEROM 9060x firmware  

11040 

FIL‐230 Titanium sintered filter, 30um  

11041 

FIL‐215 Titanium sintered filter, 15um  

14123 

Filter – analyzer bulkhead