8.2.8 HEATER SSR FAULT CORRECTION 

In the event of one of the heater solid state relays (SSR) failing in single probe mode, the heater drive 
output can be swapped to the HEATER 2 output. Select “Heater 1<‐> Heater 2”.  
If you are running in dual probe mode but not using the calibration gas or purge solenoid outputs the 
solenoid outputs can be swapped to the 2 heater outputs. Select “Heaters <‐> CalPurge”. The default 
setting is “Normal”. 

8.2.9 SSR FAIL PROTECTION 

The transmitter continuously monitors the high voltage outputs on the solid state relays that power 
the probe and purge/cal solenoids. If it detects that any solenoids are not switching off correctly then 
it mechanically isolates all high voltage outputs in order to protect the instrument. This feature can be 
switched off by setting the SSR fail protection to disabled.