8.2 CALIBRATION MODE FUNCTIONS 

8.2.1 REFERENCE VOLTAGES 

The calibration of the analogue inputs is based on the voltage of a temperature compensated voltage 
reference  integrated  circuit.  There  are  4  voltages  generated  from  the  standard  reference  voltage. 
They will vary by about 1% from one transmitter to another but can be trimmed by setting the actual 

voltages  into  CALIBRATION  menu  functions  #1  to  4.  These  voltages  should  be  measured  and  the 

reference voltage entries in the menu checked every year. 

8.2.2 OUTPUT CHANNEL 1 AND 2 CALIBRATION 

The  output  4‐20mA  calibration  can  be  set  either  AUTOMATICALLY  (default)  or  MANUALLY.  If  Auto 
Calibrated is selected the transmitter will go through an output calibration cycle when the power is 
turned on or when the AUTO CAL button is pressed in SETUP mode. This will divert the outputs back 
to the input and automatically set the 4mA and 20mA calibration.  
If  Manually  Calibrated  is  selected  then  the  4mA  and  20mA  calibration  must  be  set  in  the  next  two 
functions. For the full explanation see chapter 10.1.2. 

8.2.3 AMBIENT TEMPERATURE CALIBRATION 

The  ambient  temperature  measurement  is  used  as  the  cold  junction  temperature  for  the 
thermocouple measurements. Use an independent temperature sensor to measure the temperature 
inside  the  transmitter  case  near  the  screw  terminal  #1.  Enter  this  temperature  into  the  calibration 
function 11 by using the OPTION keys and then the ENTER key. 

8.2.4 LOW OXYGEN CALIBRATION 

The  low  oxygen  calibration  factors  can  be  used  to  fine  tune  the  oxygen  calculation  at  low  oxygen 
readings. It will not affect the measurement at 20.9%. They are included to allow oxygen probes made 
by other manufacturers to be used on a  Teledyne Analytical Instruments transmitter. The default is 
100% but can be set to between 80 and 120%. 

8.2.5 TRANSMITTER OUTPUT SCALE 

The transmitter outputs can be configured to either 4‐20mA or 0‐20mA. The two scaled outputs are 
capable  of  outputting  0‐24mA  and  will  transmit  over‐scale  unless  the  over‐scale  limit  option  is 
selected. 

8.2.6 TRANSMITTER OUTPUT LIMITING 

If the oxygen reading is invalid the output can be made to be set to 0mA, 4mA or 20mA. An invalid 
reading is when the probe temperature is below 650°C (1200°F) or the probe thermocouple has been 
detected  as  open  circuit.  The  default  is  to  set  the  output  to  20mA  but  the  function  can  also  be 

disabled. 

8.2.7 MAINS VOLTAGE DETECTION 

The default setting for the 9060Z transmitter is that it will automatically detect the mains voltage and 
frequency  in  order  to  correctly  drive  the  probe  heaters.  If  there  is any  uncertainty  in  this  detection 
system the voltage can be forced to 220‐240 or 100‐127, and the frequency can be set to 50 or 60Hz.