Page 40

background image

Model 9060Z Manual 

 

40 

 

 

 

All of the items in this list are selected as the default setting. Any of these items can be disabled from 
the common alarm by pressing the ENTER key, and they will then appear on the list in the other alarm 
relays. 

7.2.22 Operation of the Alarm Relays when an Alarm is Accepted 

An option is available to in the commissioning menu to change the operation of the alarm relay when 
an alarm has been accepted. The states for the alarm relay contacts are shown in the following table – 

Alarm state 

Accepted Relay hold setting‐

Enabled 

Accepted Relay hold setting‐

Disabled 

No alarm condition 

Closed circuit 

Closed circuit 

New alarm 

Open circuit 

Open circuit 

All alarms accepted 

Open circuit 

Closed circuit 

Alarms self cleared 

Closed circuit 

Closed circuit 

The default setting is ‘Enabled’ after a cold start.  

7.2.23 Selecting the Correct Fuel 

The 9060Z transmitter can calculate several measurements that are based on the oxygen reading and 
the temperature of the process gas. These calculations also require some details about the chemical 
makeup of the fuel. The additional calculations are: 

Oxygen deficiency  

The volume of oxygen required to bring the combustion back to 
stoichiometry  

Combustibles  

This theoretical scale is equivalent to a combustibles analyzer  

Carbon dioxide 
(maximum dry)  

The calculated maximum level of CO2 in the combustion gas  

Combustion 
efficiency  

Requires a stack thermocouple and so is only available in single probe 
mode  

Select the required fuel in function 50 by using the OPTION keys and then press the ENTER key. The 
table of values in the next 6 functions will be updated for that fuel. 

7.2.24 Reference Air Pump Options 

Normally  the  reference  air  is  supplied  from  the  transmitter  using  the  internal  pump.  The  default 
option is 5.00v. 

 

Function  

Options  

57 

Reference Air Pump  

External or Internal  

58 

Internal Pump Voltage  

Pump voltage setting, between 2.5 and 5.00 in 0.25v steps 

59 

Reference Air RH%  

If external is selected, set the RH level. (5% if instrument 
air is used)