Model 9060Z Manual 

 

15 

 

 

 

The oxygen deficiency output can be used in the same way as a combustibles analyzer to signal the 
extent of reducing conditions of combustion processes. 

4. INSTALLATION AND COMMISSIONING 

4.1 MOUNTING THE TRANSMITTER 

Surface mount the transmitter case on to a flat surface or bracket, using the four mounting brackets 
provided.  The  transmitter  should  never  be  mounted  so  that  it  is  directly  expose  to  the  sun  or  rain. 
Always leave at least 10cm of clearance around the four sides of the case. 

Make sure the temperature of the case is below 55°C, and that the radiated heat from furnaces and 
boilers is kept to a minimum. There should be adequate ventilation to maintain ambient temperature. 

Install the cables through cable glands. There are 4 holes cut in the base of the transmitter case; 2x 
17mm  &  2x  21mm.  Use  a  sharp  knife  to  cut  away  the  covering  film  for  only  the  glands  that  are 
needed.  

 

 

 

 

 

 

 

 

 

 

 

 

NOTE: ALWAYS LEAVE THE UNUSED GLAND HOLES SEALED. 

Ensure that the electrical connection complies with the local electrical requirements. (see chapter 4.6,  
Electrical Connections) 
 

 

4.2 INSTALLING A 9060H OXYGEN PROBE 

Weld a BSP or NPT socket to the flue in a suitable position for flue gas sensing. For the correct size of 
socket refer to probe data in chapter 3.1.3. The closer to the source of combustion the smaller will be 
sensing lag time, allowing better control. 

The probe has a typical response time of less than four seconds, so most of the delay time is normally 
the transit time of the gas from the point of combustion to the point of sensing.