Model 9060z manual – Teledyne 9060Z - Zirconium oxide flue gas oxygen analyzer User Manual

Page 22

background image

Model 9060Z Manual 

 

22 

 

 

 

4.8 CONNECTING THE AUXILIARY THERMOCOUPLE (OPTIONAL) 

For  9060H  heated  probes,  the  auxiliary  thermocouple  must  be  a  separate  TC  with  the  junction 
isolated from earth, mounted near to and upstream of the oxygen probe. It can be either a K, J, R or S 
type  thermocouple.  It  is  optional.  If  the  auxiliary  temperature  or  combustion  efficiency  is  not  to  be 
displayed or transmitted, then an auxiliary thermocouple is not necessary. The thermocouple should 
be connected to terminals 7 & 8. These terminals will not be available for an auxiliary thermocouple if 
a second probe has been installed. 

4.9 CONNECTING THE OUTPUT CHANNELS  

The  two  4  to  20  mA  DC  output  channels  are  capable  of  driving  into  a  1000Ω  load.  The  output 
terminals are 12 & 13 for channel 1, and 14 & 15 for channel 2. 

4.10 CONNECTING THE ALARMS 

A  common  alarm,  which  should  be  connected  for  all  installations  initiates  on  alarm  functions 
described  below.  Three  additional  alarm  relays  are  available  for  selectable  functions  as  listed  in 
chapter 9.2. Each relay has normally open contacts. The contacts will open in alarm condition. Relays 
are connected as follows: 

Relay Common Alarm  

Terminal Numbers 26 & 27  

Alarm 1  

28 & 29  

Alarm 2  

30 & 31  

Alarm 3  

32 & 33  

Common Alarms All of the following conditions will cause a common alarm ‐ 
Probe 1 (2) Heater Fail  

Ref Air Pump Fail  

Heater 1 (2) SSR Failure  

Probe 1 (2) High Impedance  

ADC Calibration Fail  

Heater SSR Leakage  

Probe 1 (2) TC Open Circuit  

Alarm Log Fail  

Probe 1 (2) Filter Blocked  

Aux TC Open Circuit  

Output 1 (2) Failure  

Gas 1 (2) Calibration Error 

 

 

BBRAM Fail  

The  watchdog  timer  is  a  special  alarm.  There  will  not  be  an  alarm  message  displayed,  but  the 
transmitter will reset. 

Alarms can be accepted by pressing the alarm button (viewing the alarm messages). 

Alarm relay 2 to 4 Select any one or all of the following for each relay. Refer chapter 9. 
In  addition,  any  of  the  selections  that  are  removed  from  the  common  alarm  relay  can  be  added  to 
relays two to four. 
High oxygen 1 (2) 
Low oxygen 1 (2) 
Very low oxygen 1 (2) 
Oxygen deviation 

Probe 1 (2) temperature low 
Calibration check 1 (2) in progress 
Probe 1 (2) purge in progress 

4.11 CONNECTING THE AUTOMATIC PURGE AND CALIBRATION CHECK SYSTEM 

CAUTION 

The  purge  and  calibration  solenoid  valves  are  supplied  with  mains  voltage.  This  supply  has 
electrical shock danger to maintenance personnel. Always isolate the transmitter before working 
with the purge and calibration solenoid valves.