7.2 COMMISSIONING MODE FUNCTIONS 

7.2.1 Internal Date / Time 

The  date  and  time  are  used  in  the  transmitter  to  run  time  based  operations  such  as  impedance 
readings  and  gas  calibration  checks.  The  correct  date  and  time  should  be  entered  by  pressing  and 
holding the OPTION UP & DOWN keys. The date and time will change faster the longer the OPTION 
key is held. 

7.2.2 Service Date 

The  service  date  can  be  used  to  keep  a  record  of  when  a  probe  was  changed.  The  RUNTIME  timer 
which keeps a track of the hours and minutes that the transmitter and probe has been operating is 
reset when the service date is changed. 
The service date and the runtime timer can be displayed as lower line items in RUN mode. 

7.2.3 Number of Probes 

The  9060Z  transmitter can  operate  in  single  or  dual  probe  mode.  By  selecting  single  probe,  all  dual 
probe options and alarms will be disabled and second probe options hidden from the user. 
If  you  are  using  the  transmitter  as  a  single  probe  device  you  should  set  the  number  of  probes  to 
‘single probe’ to disable all probe 2 alarms and display items. 

7.2.4 Probe 1 & 2 Type 

The  probe  type  function  allows  a  selection  between  a  heated  probe  and  an  unheated  probe  types. 
Heater control and alarms will be disabled if unheated probe type is selected. 

7.2.5 Probe 1 & 2 Thermocouple Type 

The  thermocouple  type  can  be  set  to  K,  J,  R  or  S.  The  thermocouple  in  the  Teledyne  Analytical 
Instruments 9060H heated probes are always K type. 9060UL/UH unheated probes are available in R 
type or S type. Other thermocouple options are made available for special installations. 

7.2.6 Auxiliary Thermocouple Type 

When the transmitter is operating as a single probe mode, the probe 2 thermocouple input (terminals 
7&8) is used as an auxiliary thermocouple input. In dual probe mode this option is unavailable. 
Use of an auxiliary thermocouple allows for basic monitoring of temperature as well as enabling the 
calculation of combustion efficiency. The thermocouple type can be set to K, J, R or S. 

7.2.7 Transmitter Output Channel 1 & 2 

The  9060Z  transmitter  has  two  fully  configurable  4‐20mA  analogue  outputs.  The  channels  can  be 
configured independently to output one of several calculated values.