Model 9060Z Manual 

 

67 

 

 

 

 

 
 

0x41 Special Instruction Function 
This command allows you to interact with the transmitter allowing access to functionality that would 
otherwise only be accessible via the keypad. The request length will vary from one command to the 
next  depending  on  what  arguments  are  required.  Requests  are  formed  the  same  as  other  modbus 
requests: 

1 byte  

1 byte  

1 byte  

0 ‐4 bytes  

2 bytes  

slave address  

command code  

function code  arguments  

checksum  

 
 

Function Code   Command  

Arguments  

0x01  

Accept All Active Alarms  

none  

0x02  

remote set key down  

1 byte keymask  

0x03  

remote reset key down  

1 byte keymask  

0x04  

Initiate Probe Impedance Check  none  

0x05  

Set Internal Clock  

4 byte rtc*  

* date expressed as seconds since 1‐Jan 2004